Reusable Ground Strap Electrode for Natus systems comes with a 150cm cable and DIN 1.5mm touch-proof (TP) connector. Ground strap is a wet type strap. Different strap lengths available. One Reusable Ground Strap Electrode per pack.

This Reusable Ground Strap Electrode is designed for use with Natus systems and can also be used with any other biosignal amplifier (EMG systems) or electrical stimulator.

The strap itself is green. It features a press-stud connector for connecting a connection cable. One 150cm connection cable with press-stud connector and a DIN 1.5mm touch-proof (TP) socket (female connector) is included in the package. This ground strap should be soaked in a saline solution before use. Ground straps are available in various lengths whereby the length refers to the contact surface of the strap and not the entire strap:

  • 10cm length x 1.5cm width (Natus part number: 9013S0714)
  • 18.5cm length x 2.0cm width (Natus part number: 9013S0735)
  • 45cm length x 2.0cm width (Natus part number: 9013S0775)
  • 75cm length x 2.0cm width (Natus part number: 9013S0745)

 

This Reusable Ground Strap Electrode has a Velcro fastener for easy mounting to the subject's wrist, ankle, or other desired location. During application, the operator should ensure that the contact surface makes contact with the subject's skin. After application, it can be let to dry and reused thereafter.

One Reusable Ground Strap Electrode per pack.

 

Technical Specifications

  • Strap Length: 10cm (4"), 18.5cm (7"), 45cm (17.7") or 75cm (29.5")
  • Strap Width: 15mm for the 10cm strap; 20mm for the other straps
  • Cable Length: 1.5m (59”)
  • Connector: DIN 1.5mm touch-proof (TP) female connector (socket)
  • Compatibility: Most EMG systems
  • Packaging: 1 x Reusable Ground Strap Electrode (Natus), Wet Type per pack
  • Package includes: Connection cable

 

Direction for use

  1. Remove the ground strap electrode from the packaging.
  2. Prepare saline solution such as salt water or electrolyte (conductive) gel.
  3. Soak the contact surface of the ground strap electrode in a saline solution. Do not soak the entire ground strap electrode but leave 2cm of felt pad near the connection cable dry!
  4. Wait 2-3 minutes for the strap to soak with the saline solution.
  5. Remove the ground strap electrode from the saline solution.
  6. Strap the ground strap electrode to the subject's wrist, ankle or other desired location.
  7. Remove the ground strap electrode after its application and let it dry.